Flow-through turbidity analyzer


The LQA5100 turbidity analyzer adopts an integrated modular design and meets the US EPA180.1 or ISO7027 standards. Unique design of water vapor elimination system, desiccant can be replaced, cleaning calibration is very easy.
Introduction Parameters Documents
  • The LQA5100 turbidity analyzer adopts an integrated modular design and meets the US EPA180.1 or ISO7027 standards. Unique design of water vapor elimination system, desiccant can be replaced, cleaning calibration is very easy.

    Measuring principle:

    Turbidity is the performance of suspended matter in water on the light through the degree of obstruction, it includes the suspended matter of light scattering and solute molecules of light absorption. The turbidity of water is not only related to the content of suspended substances in water, but also to their size, shape and refractive index. Turbidity is usually suitable for the determination of natural water, drinking water and some industrial water quality.

    The turbidimeter uses 90 ° scattering light method. After the white light sent by the light source is absorbed, reflected and scattered by the measured object during transmission, a part of the transmitted light can be irradiated on the detector in the direction of 180 °, and a part of the scattered light can be irradiated on the detector in the direction of 90 °. The light intensity received on the 180 ° and 90 ° direction detector has a certain relationship with the turbidity of the measured sewage, so the turbidity of the water quality can be calculated by measuring the intensity of the transmitted light or scattered light.

    Application occasions:

    Widely used in municipal tap water, industrial desalted water preparation, RO reverse osmosis water monitoring and other occasions.

    Product features:

    1. Built-in bubble suppression system, effectively eliminate the influence of bubbles without delaying the response time

    2. Desiccant and internal fan eliminate water vapor and measure more accurately

    3. The optical components are not in direct contact with the sample, eliminating hidden troubles

    4. Small capacity sampling cavity, improve response time

    5. Modular design, easy maintenance

    6. Convenient calibration, fast calibration, online calibration

    7. White light measurement, USEPA approved measurement method

  • Model: LQA5100
    Measurement range 0~1000 NTU automatic adjustment (0.0001~10;10 ~ 1000NTU)
    Resolution 0.0001, adjustable
    Accuracy ± 2% or ± 0.02NTU of readings below 40NTU; ± 5% of readings above 40NTU
    Response time 5~500s adjustable
    Light source White light-more than seven years of service life
    Display LCD multi-line, display with backlight
    Alarm output 2 programmable relays, 120 ~ 240VAC/2A
    Communication Interface RS485, support Modbus
    Analog output 4 ~ 20mA output, maximum load 600Ω
    water sample pressure 1~200Psi
    Flow rate 1 ml/min~1L/min.
    Sample temperature 0~50℃
    Material of liquid contact Silicate glass, silica gel, polypropylene, nylon, stainless steel
    Power 100~240VAC,80VA
    Insulation class Double-layer insulation, anti-pollution level 2
    Protection level IP66/NOME 4X
    Cleaning function Ultrasonic automatic cleaning (optional)
    Use environment Recommended for indoor use, maximum altitude 2000m, maximum humidity 95%
    Installation method Wall-mounted installation
    Dimensions 35cm × 30cm × 30cm
    Weight 2.5kg